Invacare MicroAir MA60 User Manual Page 19

  • Download
  • Add to my manuals
  • Print
  • Page
    / 24
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 18
SECTION 2—OPERATION
Part No 1148137 19 MA60 Series
9. Verify that the patient has not bottomed out by performing the following steps:
A. Ensure that the patient is lying flat on his/her back in the middle of the mattress.
B. Place four fingers between the air cushions directly underneath the sacral region
of the patient’s body.
C. Ensure that there is 3 to 4-finger width clearance between the bottom of the patient
and the mattress surface.
D. Adjust the comfort setting, if needed.
E. Wait five minutes for the mattress pressure to stabilize.
F. Repeat STEPS A-E until patient has not bottomed out and patient comfort is
achieved.
10. If the patient feels that the bed is too soft/hard, press the Soft or Firm button to adjust
the comfort settings.
11. Use a regular pillow to help support and stabilize the patient's head.
Transferring Patient From/To a Gurney
WARNING
ALWAYS engage the wheel locks of the bed and the wheel locks of the gurney
before transferring the patient between the bed and the gurney.
1. Engage the wheel locks of the bed. Refer to the owners manual provided with the bed.
2. Engage the wheel locks of the gurney. Refer to the owners manual provided with the
gurney.
3. Press the Max Inflate button to achieve maximum mattress pressure.
4. Raise or lower the bed to match the gurney height. Refer to the owners manual
provided with the bed.
5. When the mattress has reached maximum firmness, perform one of the following:
Bed to Gurney Transfer - Slide the patient onto the gurney.
Gurney to Bed Transfer - Slide the patient onto the bed.
Page view 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24

Comments to this Manuals

No comments