SECTION 7—OPERATION
Part No 1024492 27 Portable Patient Lift and Sling
FIGURE 7.1 Operating the Patient Lift - Opening the Legs of the Adjustable Base
Raising/Lowering Model Hydraulic Lifts
NOTE: For this procedure, refer to FIGURE 7.2 on page 28.
There are two controls on the pump assembly:
1. The control valve
2. The pump handle
Raising the Lift
The control valve MUST be in the closed position (control valve positioned towards pump
handle) to move the pump handle up and down to elevate the boom and the patient.
Base Closed
Base Open
Shifter
Handle
Shifter
Handle
Locked
DETAIL “B”
DETAIL “A”
Step 1
Step 2
Step 3