SECTION 7—WHEELS
Part No 1114850 45 Recumbent Handcycle Series
Adjusting Toe In/Toe Out
1. Loosen,butDONOTremovethesocketscrewsandclampsthatsecurecamberinserts
tothecamberbar(DetailʺAʺ).
2. Slowlyrotatethecamberinsertuntiltherearwheelsareapproximatelyinastraight
line.
3. Securelytightenthesocketscrewsandclampsthatsecurethecamberinsertstothe
camberbar.
4. Measurethedistancebetweenthecenterlinesattherearandfrontoftherearwheels
atapproximately12inchesfromtheground/floor.RefertoDeterminingToeIn/Toe
Outonpage 44.
5. RepeatSTEPS14untilthetoein/toeoutmeasurementislessthan1/8inch(for
maximumrollability).
Replacing Camber Inserts
NOTE:Forthisprocedure,refertoFIGURE 7.4.
1. Loosen,butDONOTremovethesocketscrewsandclampsthatsecurethecamber
insertstothecamberbar.
2. Removetheexistingcamberinsertfromthecamberbar.
3. Installthenewcamberinsertintothecamberbar.
4. Adjustthetoein/toeoutofthehandcycle.RefertoDetermining/AdjustingToeIn/Toe
Outonpage 44.
FIGURE 7.4 - Replacing Camber Inserts
Clamp
Camber Bar
Socket Screw
Camber
Insert
Frame
Page view 44